Abstract
The description is given of a demountable, continuously evacuated experimental X-ray tube with a transmission target, in which modifications in design of anode, filament, and filament hood can be made. The pumping plant, vacuum technique and electrical circuits are described. Operating at 35 kV to 40 kV and 12 mA with a focal skin distance of 5 cms and a circular field of 5 cm diameter, the tube has an output of 200 r. per minute. It is held in an earthed insulator, a photograph of which is given.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.