Abstract

Compare the difference of the results referred to the WHO standard rabies immunoglobulin and the national standard human rabies immunoglobulin used in the rapid fluorescent focus inhibition test (RFFIT). Setting the WHO standard immunoglobulin and the national standard immunoglobulin in the same system and testing 12 human serum at the same time. Compare the fluorescence percentage of the two different standard immunoglobulin; compare the 12 serum results calculated from the two different standard immunoglobulin used the calculation formula of neutralization antibody titer. The Results display that the 50% percent of the two standard immunoglobulin are all between the fifth and the sixth well, but the percentage of the national standard immunoglobulin is lower than the WHO one. The same testserum result calculated from the WHO standard immunoglobulin is little higher than the national one. There is difference in the WHO standard immunoglobulin and the national one, but there is no influence in the results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.