Abstract

Flattening filter free linear accelerator (FFF LINAC) has been installed in Indonesia. To ensure the accuracy of FFF irradiation, we evaluate the characteristic of FFF in regular and small fields. We employed the Monte Carlo (MC) simulation of FFF LINAC 6 MV as a standard reference in this study. Then, we compared the calculation of FFF beam at central and lateral axis to the measurement and TPS at 10 × 10, 1 × 1, 2 × 2, 3 × 3, 4 × 4 cm2 field sizes. Output factor (OF) and beam quality such that TPR20,10 and penumbra of FFF LINAC were evaluated. TPR20,10(S) for 10 × 10 cm2 was 0.01 differ between MC and measurement whereas it varied from 0.588 to 0.703 for small fields. Dose difference of lateral axis was agreed within 3% except for the penumbra region. Output factor of small fields measurement indicated that field size of 1 × 1 cm2 had a large discrepancy to MC according to this works. The TPR20,10(10) of MC, TPS, and measurement was not a significant difference, while the OF was tending to a large deviation in 1 × 1 cm2. The results showed that our FFF LINAC could be used for small field until 2 × 2 cm2.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.