Abstract
On the basis of discussing the feasibility of gradient matching layer design, the simulation of single matching layer, double matching layer and triple matching layer is carried out by PiezoCAD software. The simulation results indicate that the energy utilization efficiency of ultrasonic transducers will increase with the increase of acoustic matching layers and the −6dB bandwidth. Based don simulation, the gradient acoustic matching layer is prepared, and the experiment results prove that the ultrasonic transducer prepared contains gradient acoustic matching layer has a wider bandwidth. The feasibility of the ultrasonic transducer with gradient impedance and the influence of the matched layer material on the acoustic attenuation coefficient is also discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.