Abstract
We compared the frequency of a binary endocardial appearance in patients with hypertrophic cardiomyopathy (HCM) and Anderson-Fabry disease (AFD). A recent study suggested that a binary endocardial appearance is a highly sensitive and specific discriminator of AFD from other causes of hypertrophic cardiomyopathy (HCM). Fourteen patients with AFD (55.4 +/- 9.9 years, 9 men) and 14 patients with HCM (57.2 +/- 10.9 years, 9 men) were randomly selected from a dedicated patient database. Two-dimensional echo images were blindly reviewed by 2 experienced echocardiographers. Maximum left ventricular (LV) wall thickness, LV end-systolic dimension, fractional shortening, and left atrial size were similar in the 2 patient groups. The LV end-diastolic dimension was smaller in patients with HCM (p = 0.04). A binary sign was present in 8 of 28 patients (29%). The sensitivity and specificity of the binary sign as a discriminator of AFD from HCM were 35% and 79%, respectively. A binary sign was present in only 1 patient with LV wall thickness <15 mm. The binary endocardial appearance lacks sufficient sensitivity and specificity to be used as an echocardiographic screening tool.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.