Abstract

PurposeAn artificial placenta (AP) utilizing extracorporeal life support (ECLS) could avoid the harm of mechanical ventilation (MV) while allowing the lungs to develop. MethodsAP lambs (n = 5) were delivered at 118 days gestational age (GA; term = 145 days) and placed on venovenous ECLS (VV-ECLS) with jugular drainage and umbilical vein reinfusion. Lungs remained fluid-filled. After 10 days, lambs were ventilated. MV control lambs were delivered at 118 (“early MV”; n = 5) or 128 days (“late MV”; n = 5), and ventilated. Compliance and oxygenation index (OI) were calculated. After sacrifice, lungs were procured and H&E-stained slides scored for lung injury. Slides were also immunostained for PDGFR-α and α-actin; alveolar development was quantified by the area fraction of alveolar septal tips staining double-positive for both markers. ResultsCompliance of AP lambs was 2.79 ± 0.81 Cdyn compared to 0.83 ± 0.19 and 3.04 ± 0.99 for early and late MV, respectively. OI in AP lambs was lower than early MV lambs (6.20 ± 2.10 vs. 36.8 ± 16.8) and lung injury lower as well (1.8 ± 1.6 vs. 6.0 ± 1.2). Double-positive area fractions were higher in AP lambs (0.012 ± 0.003) than early (0.003 ± 0.0005) and late (0.004 ± 0.002) MV controls. ConclusionsLung development continues and lungs are protected from injury during AP support relative to mechanical ventilation. Level of evidencen/a (basic/translational science).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call