Abstract

The overall recovery of the separated113mCd with added Cd carrier from 50 l samples of natural water was determined by a secondarysource tube excited energy dispersive X-ray fluorescence spectrometer measuring the amount of cadmium in the final CdS+CuS coprecipitate on a 25 mm diameter Millipore filter. The method enables to determine 1–20 μg Cd in 100 sec. with a limit of detection of 0.8 μg.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.