Abstract
Methods of flow visualization using oil streaking are established techniques for investigating surface shear and near wall flow patterns. Recent studies have used an array of oil dots on a surface which form streaks when exposed to shear forces. This method is generally qualitative, but it is possible to make quantitative measurements of the shear if the oil streaks have been calibrated. This paper presents the application of a quantitative oil streak method to the HeartQuest left ventricular assist device (LVAD). An array of dots was applied to the top housing of the pump, yielding quantitative values for the shear and qualitative patterns of the near wall flow in that region. The results were used to locate regions likely to promote thrombosis, such as stagnation points or recirculation regions. Regions of high shear, where hemolysis might occur, also can be identified with this method. In addition to being an important design technique, quantitative oil streaking assisted in the verification of computational fluid dynamics results within the HeartQuest LVAD.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.