Abstract

To assess the suitability of the i-STAT portable analyser for use by non-laboratory personnel, we measured blood gases and pH in venous blood samples from 100 dogs. Deming's regression and bias plots were used to compare i-STAT results with those obtained by laboratory professionals using two different autocalibrated benchtop analysers. Overall accuracy of the portable analyser proved excellent for pH, pO 2, and pCO 2 ( r=0.978, 0.968 and 0.997, respectively), with Deming's regression slopes close to 1.00 (0.96, 0.97 and 1.08 for pH, pO 2, and pCO 2, respectively) and intercepts close to zero (0.28, 0.47 kPa and 0.46 kPa for pH, pO 2, and pCO 2, respectively). The accuracy of the i-STAT was also satisfactory for calculated parameters: bicarbonates, total CO 2, base excess and oxygen saturation. Our findings show this portable analyser to be a valid substitute for expensive benchtop analysers in situations requiring mobility, or when small numbers of tests are to be performed by users not specialized in laboratory techniques.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.