Abstract

A micro-Faraday array detector (row of miniature 32 thin film strip electrodes wire bonded to an on-chip integrated capacitance transimpedance amplifier) was tested for use in a multiple collector secondary ion mass spectrometry. The detector was mounted on a standard IMS3f SIMS instrument in place of a microchannel plate. The measurements were performed by using a silicon sample bombarded by either or Cs+ primary ions with or without D2O flooding and detecting secondary ions of Si±, SiH±, O-, SiD−, OD−. A parallel detection of near masses of 29Si and 28SiH or 18O and 16OD was demonstrated at a sensitivity level ∼230 counts/s.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.