Abstract
Temporary right heart bypass has shown to improve hemodynamic stability and safety in beating heart revascularization. We sought to evaluate feasibility and safety of a right ventricular assist device percutaneously implanted in the right jugular vein. The A-Med jugular coaxial cannula (A-Med Systems Inc, West Sacramento, CA) is designed for percutaneous implantation. Blood from the right heart is drained through the outer tube of this two-cannula device to a microcentrifugal pump and returned into the pulmonary artery through the inner tube. In 10 patients scheduled for elective coronary bypass grafting without cardiopulmonary bypass, a total of 27 coronary anastomoses were performed with right heart support. Arterial pressure was significantly higher with right heart support when the heart was dislocated to access posterior and posterolateral anastomosis sites. Implantation through the right internal jugular vein was feasible without complications in all patients and facilitated the procedure. This initial study suggests safety and feasibility of temporary right heart support using percutaneous jugular access for posterior and posterolateral coronary bypass grafting.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.