Abstract
A new temperature controlled transfer standard package for absolute and differential pressure transducers has been developed at the Ulusal Metroloji Enstitusu (UME). Due to direct (metal-metal) touching of the temperature controlled plate of the temperature controlled unit with the pressure sensor housing the temperature influence of electronics on the pressure sensor is ruled out. The pressure sensor temperature inside of cast aluminum housing can be controlled at any fixed temperature in the temperature range of 15 °C − 30 °C with stability ± 2 mK. The mentioned transfer standard package minimizes the ambient temperature effects upon the sensor, denies the need to apply a correction for thermal transpiration effect and thereby increases accuracy of the transducer. The transfer standard packages have demonstrated excellent short-term zero stability which does not exceed ± 5 mPa and a maximum drift rate of 0.03 mPa h−1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.