The temperature of the inspired humidified gases of an intubated patient may be monitored accurately, close to the airway, by a commercially available thermistor probe and alarm/cut-out device for use with heated water humidifiers. The apparatus is designed to alarm when the inspired gas temperature exceeds the pre-set temperature and when the thermistor probe is in open or short circuit. When the temperature alarm is activated, the humidifier heater is switched off automatically, thus preventing prolonged transmission of excessively hot gases to the airway. The apparatus has been used successfully in this hospital for several months, requiring minimal observer attention and maintenance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.