In this study, the effective dose rate and shielding thickness for a PET/CT facility are calculated by the analytical formula and PHITS code. The calculations at 9 different points in the unshielded and shielded rooms of the PET/CT were performed and the results were compared. The dose administered and the workload were estimated to be 15mCi and 40 patients per week. The required barrier thickness based on the weekly dose limit is 100 µSv for the controlled areas and 20 µSv for uncontrolled areas, corresponding to the 1 mSv/year limit to the general public.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.