Abstract

In this study, the systematic errors of an X-ray single-grating interferometer based on the Talbot effect were investigated in detail. Non-negligible systematic errors induced by an X-ray camera were identified and a method to eliminate the systematic error was proposed. Systematic-error-free measurements of the wavefront error produced by multilayer focusing mirrors with large numerical apertures were demonstrated at the SPring-8 Angstrom Compact free electron LAser. Consequently, wavefront aberration obtained with two different cameras was found to be consistent with an accuracy better than λ/12.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.