Abstract

Targeted analysis of sequential window acquisition of all theoretical mass spectra (SWATH-MS) requires the spectral library, which can be generated by shotgun mass spectrometry (MS) or by the pseudo-spectra files directly obtained from SWATH-MS data. The external library generated by shotgun MS is employed in most SWATH-MS research. However, performance of the internal library, which is constructed by pseudo-spectra files, in the targeted analysis of SWATH-MS has not been systemically evaluated. Here, we show that up to 40% of the peptides detected by the internal library were not overlapped with those detected by the external library for most SWATH-MS data sets. However, the internal library did not identify extra phosphopeptides compared with the external library for phosphoproteomic SWATH-MS data. Therefore, the internal library should be incorporated into the external library for targeted analysis of nonphosphoproteomic SWATH-MS, given that it can significantly increase the number of peptides of SWATH-MS without requiring additional instrument measurement time.