Abstract

Aim: To develop and validate a system that can wirelessly acquire gastric slow waves and deliver electrical pulses to the stomach. Materials & methods: The system is composed of a front-end and a back-end unit connected to a computer, which runs a custom-made graphical user interface. The system was validated on benchtop and in vivo studies. Results: Benchtop validation showed an appropriate frequency response to acquire slow waves. Moreover, the system was able to deliver electrical pulses at amplitudes up to ±10 mA. Slow wave activity recorded from the stomach of rats was in the range of approximately five cycles per minute (cpm). Pulses delivered to the stomach of a rat every 15 s and reduced the activity to 4 cpm during the stimulation period. Conclusion: This study reports the first wireless system and methodology that can be used to acquire slow waves and deliver electrical stimulation to the stomach of small freely behaving animals.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.