Abstract

This review details some peripheral nervous system (PNS) targets and electrode designs used for electrical stimulation. It investigates limitations in current knowledge of safe electrical stimulation and possible future electrode developments. Current PNS targets are large, leading to poor resolution and off-target side-effects. Most clinical devices are platinum or platinum/iridium embedded in an insulation material. Their safety is usually guided by the Shannon plot, which is not valid for the PNS. New electrode designs are needed to target smaller nerve fibers, enabling higher resolution electrical therapies with fewer off-target side-effects. Damage can occur through biological and electrochemical mechanisms. Greater mechanistic understanding is required to ensure safe and efficacious, long-term electrical stimulation with new electrode materials, geometries and stimulation waveforms.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.