Abstract
An automated technique involving switching valves and a filter assembly has been developed and evaluated for the on-line precipitation of proteins and peptides from plasma samples. In the set-up, the proteins were precipitated on-line by injecting the plasma sample into a stream of organic precipitating agent. The precipitates so formed are filtered on-line by a set of filter assemblies consisting of ordinary in-line HPLC solvent filters. Evaluation of the technique was performed using ibuprofen and a mixture of three estrogens, estradiol, equilin and estrone, spiked in dog plasma. The coefficients of variation (C.V.) for system suitability parameters were below 10%. Absolute recovery of ibuprofen in plasma ranged from 80% for 100 μg/ml to 114% for 5 μg/ml spiked concentrations, respectively. Resolution for equilin and estrone, two closely eluting peaks, was 1.79 (C.V. = 5.8%, n = 7). The switching-valve—filter assembly had no significant effect on the efficiency of the HPLC system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.