Abstract

Matrix gradient coils with up to 84 coil elements were recently introduced for magnetic resonance imaging. Ideally, each element is driven by a dedicated amplifier, which may be technically and financially infeasible. Instead, several elements can be connected in series (called a “cluster”) and driven by a single amplifier. In previous works, a set of clusters, called a “configuration,” was sought to approximate a target field shape. Because a magnetic resonance pulse sequence requires several distinct field shapes, a mechanism to switch between configurations is needed. This can be achieved by a hypothetical switching circuit connecting all terminals of all elements with each other and with the amplifiers. For a predefined set of configurations, a switching circuit can be designed to require only a limited amount of switches. Here we introduce an algorithm to minimize the number of switches without affecting the ability of the configurations to accurately create the desired fields. The problem is modeled using graph theory and split into 2 sequential combinatorial optimization problems that are solved using simulated annealing. For the investigated cases, the results show that compared to unoptimized switching circuits, the reduction of switches in optimized circuits ranges from 8% to up to 44% (average of 31%). This substantial reduction is achieved without impeding circuit functionality. This study shows how technical effort associated with implementation and operation of a matrix gradient coil is related to different hardware setups and how to reduce this effort.

Highlights

  • Matrix gradient coils were introduced for both magnetic resonance image acquisition and B0 shimming [1,2,3,4,5,6]

  • Coil elements of a matrix gradient coil are smaller than typical dimensions of conventional gradient coils, which allows for faster switching of currents

  • We have shown how such a configuration can be optimized by assigning coil elements to clusters such that a desired target field shape can be approximated with high accuracy

Read more

Summary

Introduction

Matrix gradient coils ( termed multi-coils or multicoil arrays) were introduced for both magnetic resonance image acquisition and B0 shimming [1,2,3,4,5,6]. Almost 2 decades earlier, a similar concept called matrix shim coil system was introduced for nuclear magnetic resonance [7]. Such coils consist of a multitude of (up to 84) compact coil elements, which are in the most trivial configuration individually supplied by a current amplifier. With matrix gradient coils, encoding fields can be adapted to the actual FOV needed for the given imaging application. This may potentially be more accurate and/or more efficient [4], for small FOVs

Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call