Abstract
A simple and effective ultra-high-performance liquid chromatography assay linked to tandem mass spectrometry (UHPLC-MS/MS) for measuring cortisol and cortisone levels in human sweat has been developed and validated. A noninvasive world standard sweat collecting equipment was utilized to collect samples. The samples were analyzed using an Atlantis dC18 (2.1 × 100 mm, 3 μm) column with a 2 mM ammonium acetate and acetonitrile (1 : 1, v : v) mobile phase. In an isocratic condition, the mobile phase was delivered at a flow rate of 0.3 ml/minute. A positive electrospray ionization interface with multiple-reaction monitoring mode was used to provide simultaneous quantification of cortisol, cortisone, and internal standard at transitions of 363.11 to 121.00, 361.18 to 163.11, and 367.19 to 121.24, respectively. The method was validated for cortisol and cortisone determination over a concentration range of 0.5-50 ng/mL The detection limits for cortisol and cortisone in human sweat were 0.3 and 0.2 ng/ml, respectively. The interday coefficients of variation of cortisol and cortisone were ≤8.5% and ≤10.01%, whereas bias was in the range from -7.9% to 2.1% and from -4.3% to 3.0%, respectively. The assay was successfully applied to evaluate the cortisol-to-cortisone ratio in sweat samples collected from healthy adult volunteers.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have