The Neuflex silicone metacarpophalangeal (MCP) joint replacement has previously been shown to provide improved movement, particularly flexion, when compared with the Swanson implant. In this study, we reviewed the outcome of 237 Neuflex implants in 66 patients with a follow-up of up to 7 years. Kaplan-Meier analysis using revision as the end stage revealed survival at 7 years to be 88%. If however an implant fracture, as seen radiographically, is taken as the end point, the survivorship drops to 68% at 7 years. These figures are comparable with a similar analysis for the Swanson implant. We conclude that the improved range of motion of the Neuflex implant demonstrated previously does not result in either a higher or a lower revision or implant fracture rate when compared with the Swanson implant.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.