Contactless vital sign monitoring systems are becoming increasing in demand for a wide range of biomedical applications. Millimetre-wave radars and embedded signal processing are the most promising technologies to enable non-contact vital signs monitoring. In this work, the challenging task of heart rate estimation from radar data has been addressed. Three different radar systems from Infineon, Texas Instruments and Acconeer, and four algorithms, FFT, Median-FFT, STFT and Median-STFT, have been analysed and compared against a reference sensor. Accuracy, as well as power figures, have been reported for all the radar systems. A dataset of 16 volunteers has been acquired, yielding a total of 400 minutes of radar-recorded vital sign data. The accuracy of the four investigated algorithms has been reported on average and per subject for every radar. The algorithm exploiting the Short Time Fourier Transform (STFT) is able to achieve an error as low as 0.02% on a single person and of 6.4% in heart rate estimation on average across the whole dataset.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.