Abstract

Osteointegration is dependent on a variety of biomechanical and biochemical factors. One factor is the wettability of an implant surface that is directly influenced by its surface energy. This investigation used the Zisman plot to determine critical surface tension as one representative measurement of surface energy. The effects of surface treatment, bulk grain size, and surface roughness on the critical surface tension of unalloyed titanium (Ti) were examined. Radio frequency glow discharge-treated Ti had the highest critical surface tension, followed by the passivated and heat-sterilized conditions. Titanium with no surface treatment had the lowest critical surface tension. The surface energy of Ti with an average grain size of 23 microns was not significantly different from that with a grain size of 70 microns. Surface roughness was shown to cause significant difference in measurements and definitely should be considered in studies of this kind.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.