Abstract
The supralinear response of thermoluminescent dosimeters after irradiation with gamma-or beta-rays can be decomposed by use of the cummulative Poisson distribution into a sum of 1-or-more and 2-or-more hit components, suggesting the coexistence of both types of trap structures. Following the response of other 1-hit detectors, and of biological cells to energetic heavy ions, we expect and find that the supralinear response tends to disappear with alpha particle, and with neutron irradiation, and that the supralinear, or 2-hit, component tends to be more responsive to high LET (Linear Energy Transfer) radiations. Since the response of biological cells to gamma-rays is also supralinear, the supralinear component of the TLD response can be expected to have the capacity to mimic the response of biological cells and tissues to radiations of different quality.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.