Abstract

Purpose: In this study, we evaluated the effect of grid size on dose calculation accuracy using 2 head & neck and 2 prostate IMRT cases. Methods: Dose distribution of four IMRT plan data were calculated at four calculation grid sizes (1.25, 2.5, 5, and 10 mm) and the calculated dose distributions were compared with measured dose distribution using 2D diode array detector. Results: there was no obvious difference in pass rate of gamma analysis with 3 mm/3% acceptance criteria except 10 mm grid size, we found that the pass rates of 2.5, 5 and 10 mm grid size were decreased 5%, 20% and 31.53% respectively. The calculation time were about 11.5 min, 4.77 min, 2.95 min, and 11.5 min at 1.25, 2.5, 5, and 10 mm, respectively and as the grid size increased to double, the calculation time decreased about one‐half. The grid size effect was observed more clearly in the high gradient area than the low gradient area. Conclusions: In conclusion, 2.5 mm grid size is considered acceptable for most IMRT plans but at least in the high gradient area, 1.25 mm grid size is required to accurately predict the dose distribution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.