Abstract

A 72-year-old man with nonischemic cardiomyopathy was referred because his implantable cardioverter defibrillator had failed to terminate spontaneous ventricular fibrillation (VF). Defibrillation threshold (DFT) testing confirmed that 830-V shocks failed to defibrillate VF despite optimization of the biphasic waveform and reversal of shock polarity. The placement of a new right ventricular lead and the addition of a subcutaneous array failed to defibrillate VF at 830 V. The combination of a subcutaneous array and azygos vein coil successfully defibrillated VF. The mechanism for successful DFT reduction was likely greater current supplied to the posterior basal left ventricle by the azygos vein lead.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.