Abstract

A comparison of standard platinum resistance thermometers (SPRTs) has been carried out between the BIPM and the Laboratoire commun de métrologie LNE-CNAM using the melting point of gallium and the triple point of water. The temperature difference at Ga between the BIPM and the LNE-CNAM was determined as 108 µK with an associated combined standard uncertainty of 223 µK. This outcome indicates a present temperature difference of the BIPM of −65 µK with respect to an Average Reference Value from an earlier comparison, with an associated uncertainty of uc = 262 µK. It should be noted that the present BIPM scale is not independent, but traceable to cells used in KC7. The comparison results validate the re-establishment of thermometer calibrations at the BIPM for internal use.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCT, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).