Abstract

Main textThis report shows the results of a bilateral comparison of temperature standards between Instituto Nacional de Metrologia, Qualidade e Tecnologia (Inmetro, Brazil), and Trinidad and Tobago Bureau of Standards (TTBS, Trinidad and Tobago), carried out between January and July of 2020. The measurands of this comparison were the W(t90) ratios of two Standard Platinum Resistance Thermometers (SPRTs), used as transfer standards, measured at the zinc freezing point (Zn FP, 419.527 °C), tin freezing point (Sn FP, 231.928 °C), gallium melting point (Ga MP, 29.7646 °C), triple point of water (TPW, 0.01 °C) and mercury triple point (Hg TP, -38.8344 °C).To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.The final report has been peer-reviewed and approved for publication by the CCT, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.