Abstract
We report submicron imaging (∼0.75 μm resolution) of complex magnetic susceptibility in a micron-size ferromagnetic heterostructure using time-resolved scanning transmission x-ray microscopy. The real and imaginary parts of the susceptibility are extracted from the phase and amplitude of the small-angle (<20°) rotational response of the local magnetization under microwave excitation. Frequency-dependent response patterns were observed in an incompletely saturated bilayer element. The technique is extensible to higher frequencies (to ∼10 GHz), better spatial resolution, and layer-specific measurement.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.