Purpose: A research prototype for high‐performance dual‐energy (DE) imaging of the chest is under development. This paper discusses the development and characterization of a cardiac gating system designed to precisely trigger the imaging system according to cardiac phase and minimize anatomical misregistration due to heart motion. Method and Materials: A fingertip pulse oximeter was employed to measure the peripheral pulse waveform and trigger x‐ray exposures during the quiescent phase of the heart (diastole). Temporal delays accounted in the timing scheme include physiological pulse propagation, waveform processing, and imaging system delays (filter‐wheel, bucky‐grid, and flat‐panel detector). An empirical model of the diastolic period allows calculation of the implemented delay, timp, required to trigger correctly at any patient heart‐rate. Performance was evaluated in terms of accuracy and precision of diastole‐trigger coincidence and expert assessment of cardiac motion artifact in gated and ungated DE images. Results: The model suggests a triggering scheme characterized by two heart‐rate (HR) regimes: below a HR‐threshold, sufficient time exists to expose on the same heartbeat ; above the HR‐threshold, a characteristic timp(HR) delays exposure to the subsequent heartbeat, accounting for all fixed and variable system delays. Initial implementation indicated 83% accuracy in diastole‐trigger coincidence. By modifying the HR estimation method (reduced temporal smoothing of the pulse waveform), trigger accuracy of 100% was achieved. Cardiac‐gated DE patient images demonstrate significantly reduced cardiac motion as assessed by expert radiologists.Conclusion: A pulse oximeter combined with a cardiac model provides accurate x‐ray triggering and significantly reduces heart motion artifacts. A simple fingertip clip presents logistic, cost, and workflow advantages compared to ECG. The system has been implemented in a clinical research trial, with gated and ungated arms allowing characterization of the impact of cardiac motion artifact on diagnostic performance. Conflict of Interest: Research sponsored in part by Eastman Kodak.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.