Abstract
Objective: To explore the optimal matching of the current parameters about artificial facial nerve (AFN) for inducing contraction of the orbicularis oris muscle (OOM) in rabbit with peripheral facial paralysis. Methods: According to the combinations of different stimulus parameters, at different time points, we used AFN to induce contraction of the affected side OOM of the paralyzed rabbit in waking state. We recorded the current amplitudes of the threshold stimulation and peak stimulation under each combination, then compared the total charge of the stimulus consumption under different combinations. Results: Compared the total stimulus charge consumption of the AFN threshold stimulation and that in the peak stimulation respectively under different stimulus frequency and pulse width matching combinations, we found that the frequency, the pulse width and the interaction of the frequency and pulse width among different groups could affect the total charge, the differences were statistically significant (P<0.05). Conclusions: When AFN stimulus frequency is 60 Hz and the stimulus pulse width is 100 μs, the output current intensity is relatively lowest and the total consumed charge is the lowest under the premise of effectively inducing the paralytic side OOM contraction. Thus we recommend this stimulus parametric combination as the optimal combination for meeting the low power consumption of AFN.