Abstract

Objective To evaluate the pre-hospital diagnostic reliability of real-time tele-transmission of 12-lead electrocardiogram of patients with ST-segment elevated acute myocardial infarction (STEMI).Methods The 12-lead electrocardiogram was simultaneously recorded with real-time tele-transmission system and a conventional electrocardiograph in 40 STEMI cases.The width and amplitude of each wave,the deviated amplitude of ST-segment in the same leads were compared by t-test and rank-sum test.Results There were no statistical differences in the width and amplitude of P wave,QRS wave and t wave as well as the deviated altitude of ST-segment between the two separate electrocardiographs (P >0.05).There was a significant positive correlation between the two ECG devices in respect of ST-segment elevated altitude (r =0.912,P =0.000).The differential ability of ST-segment elevation between two separate ECG devices kept highly consistent (Kappa value:0.976).Conclusions Real-time tele-transmission of 12-lead electrocardiogram is reliable for the pre-hospital diagnosis of STEMI. Key words: Acute myocardial infarction; Electrocardiogram/remote-electrocardiogram; Real-time ; Tele-medicine; Remote ECG monitoring; Wireless/3G; Bluetooth ; Pre-hospital diagnosis; Reliability ; Accuracy