Abstract

In this work, personal thermoluminescence dosimeter (TLD) cards type of GN-6770 (holder type 8806) from Harshaw were used for personal neutron dosimetry. The response of the dosimeters has been determined in terms of the personal absorbed dose and personal dose equivalent for different neutron energy components, based on the recommendations of ICRP-60 and ICRU-49. Neutron irradiation was performed using a 5 mCi Am–Be neutron source. The TLD reader, type Harshaw 6600, was installed and calibrated for accurate neutron doses equivalent to gamma-ray doses. It was found that fast neutron doses measured by TLD (badges or cards) are in agreement with those measured by neutron TE (tissue equivalent gas) ionization chambers and neutron monitors. Thermal neutron doses measured by TLD cards were overestimated when compared with those measured by neutron monitors. Additional Cd was used to reduce thermal neutron doses to be in agreement with actual thermal doses. Other configurations for TLD crystals are also suggested for accurate thermal neutron dose measurements.