Abstract
The main purpose of this study is to measure the contribution of the thermal and fast neutron dose along the central axis of the 15 MV Elekta Precise linac in a tissue equivalent phantom. In order to achieve this purpose, different points were selected in three field sizes of 5×5cm2, 10×10cm2 and 15×15cm2. Fast and thermal neutrons were measured using CR-39 nuclear track detectors with and without thermal neutron converter of 10B, respectively. According to the results, the fast neutron dose equivalent was decreased as the depth increased (field size 5×5, 10×10 and 15×15cm2 fall from 0.35 to 0.15, 0.5 to 0.3 and 0.5 to 0.3, respectively). Thermal dose equivalent was increased as the depth increased in the tissue equivalent phantom (field size 5×5, 10×10 and 15×15cm2 rise from 0.1 to 0.4, 0.4 to 0.8 and 0.4 to 0.9, respectively). In conclusion, at depth <3cm, most existing neutrons are fast and CR-39 films are sensitive to fast neutrons; therefore, they are more appropriate than thermoluminescent dosemeters in measuring neutron dose equivalent.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.