Abstract

The accuracy of today's coordinate measuring machines (CMM) has reached a level at which exact knowledge of each component is required. The role of the probe tip is particularly crucial as it is in contact with the sample surface. Understanding how the probe tip wears off will help to narrow the measurement errors. In this work, wear of a sapphire sphere was studied for different scanning conditions and with different sample materials. Wear depth on the probe was investigated using an automated process in situ on the METAS micro-CMM and completed by measurements with an atomic force microscope. We often found a linear dependence between the wear depth and the scan length ranging from 0.5 to 9 nm m−1, due to variations in scan speed, contact force or sample material. In the case of steel, the wear rate is proportional to the scan speed, while for aluminum several processes seem to interact. A large amount of debris was visible after the tests. Except for aluminum, wear was visible only on the sphere and not on the sample. Sapphire/steel is the worst combination in terms of wear, whereas the combination sapphire/ceramic exhibits almost no wear.