Abstract
This paper validates a technique to add statistical noise to a Computed Radiography (CR) in order to simulate accurately how the same image would appear if taken at a reduced tube current. To that, a noise addition software has been developed to create lower dose CR using existing pediatric radiographies based on the selection of lower X-ray tube current. The effect of different milliAmpere-seconds (mAs) setting on image quality has been evaluated using the CDMAM 3.4 phantom and the obtained results show good agreements between the simulated and real images in terms of noise measurement. The new CR images allow medical researchers to study how lower dose affects the patient diagnosis.KeywordsComputed radiographynoisedose reduction
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.