Abstract

This paper validates a technique to add statistical noise to a computed radiography (CR) image in order to simulate accurately how the same image would appear if taken at a reduced tube current. To that end, a noise addition simulation software has been developed in order to create lower-dose pediatric CR selecting the desired lower X-ray tube current.The effect of different tube current settings (in milliampere·seconds) on image quality has been evaluated using the CDMAM 3.4 phantom, and the obtained results show good agreement between the simulated and real images in terms of noise measurement. The new CR images allow medical researchers to study how lower dose affects the patient diagnosis without taking new images.The developed algorithm will be used in future medical research on determining the minimum tube currents necessary for adequate diagnoses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.