Abstract

A comparative study of n-γ discrimination by the digital charge comparison method was carried out for NE213 and BC501A scintillators of different size coupled to a 130 mm diameter XP4512B photomultiplier. Particularly, the scintillators of diameter 50 mm × 50 mm size are compared with those of 4 l volume (160 mm in diameter and 200 mm in depth). The figure of merit, M, of 3.81 measured with the diameter 50 mm × 50 mm BC501A scintillator at 1 MeV energy lost by recoil electrons, is much better than M of 2.05 as observed with the 4 l volume cell. This is the effect of the reduced photoelectron yield by about 50% determined for the 4 l volume scintillator. Moreover, the slowing down of the light pulse and multiscattering of neutrons have a further influence on the lowering of the M factor. The study of the M factor versus the initial delay and width of the gate set at the slow component showed that the early light of the slow component represented often by the intermediate component with the decay time constant of about 30 ns is of great importance for the n-γ discrimination. Particularly, for the 4 l volume cells it is sufficient to limit the width of the gate at the slow component to about 300 ns. The comparison of the NE213 and BC501A scintillators showed that the BC501A scintillator exhibits a larger light yield evidently observed for the 4 l volume cell and thus giving a better n-γ discrimination.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.