Abstract

The effect of an intense magnetic field on an 1100-type capacitance thermometer (made from an SrTiO3 glass-ceramic crystallized at 1100°C) was accurately measured at five temperatures between 1.5 and 4.2 K. At each manostatically maintained temperature, the capacitance changes were measured five times between B=0 and B=14 T, and it was found that these changes did not exceed the measurement uncertainty, ±0.3 pF≅±1 mK. Similar results were obtained on a second 1100-type thermometer. Some suggestions are made for the use of this type of thermometer in magnetic field experiments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.