Abstract
Thin membrane transducer can be used for detecting the surface stress induced by the bio-molecular reactions. Designing research of the capacitive thin membrane transducer is proposed in the paper. The relationship between the sensitivity of the surface stress-based biosensors and the geometrical parameters of membrane was analyzed using the finite element method. The thermal strain mismatch resulting from the CTE difference of the two-layers of the membrane was also studied. The structure characteristics were analyzed systematically, including shape, size, gold coverage, which are normally considered in the design of the biosensors. Furthermore, a novel bi-membrane structure, which has high signal-to-noise ratio and can minimize the influence of the environment, is proposed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.