Abstract

The objective of this study was to study the dose reduction effect on occupational exposure of operators and public exposure other than operators when a backscatter shield and a rectangular collimator are used in conjunction with a handheld intraoral X-ray unit. The occupational exposure was reduced to 40% when the backscatter shield was attached to the cone-tip, to 13% when the rectangular collimator was attached and to 7.7% when the backscatter shield and rectangular collimator were used together. On the other hand, the public exposure was reduced to 20% when the rectangular collimator was attached, but the backscatter shield was not effective in reducing the public exposure. Attaching a backscatter shield is effective in reducing the occupational exposure, and a rectangular aperture is effective in reducing the occupational exposure, as well as the public exposure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.