Abstract
Potential ways to avoid limitations and improve analytical sensitivity in the determination of fluorine (F) in solid materials using wavelength dispersive X-ray fluorescence (WDXRF) spectrometry were investigated. Chemical shift was not significant in the F determination using WDXRF spectrometry, potentially because of high electronegativity of F. In a variety of matrices containing F, an absorption/enhancement effect noticeably altered the F intensity. A sample containing high-molecular-weight constituents seemed to absorb the emitted F-Kα line more than its counterparts. An additional calibration curve established using standards with similar matrices to samples is recommended. In addition, the intensity induced from the F-Kα increased by more than 9-fold by making a pressed pellet with a low dilution (9:1) method using polyvinyl alcohol and liquid binder. Peak overlap caused by the proximity between F-Kα and Fe-Lα or Mn-Lα is likely to disturb the quantitative F determination. It can be properly overcome by employing a correction factor derived from the slope of the measured X-ray intensity at the Bragg angle of 43.174° to Fe or Mn content in the sample. Consequently, WDXRF technique can be successfully used to assess the F content in solid materials by following the strategies proposed in this study.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.