Whole stones (N = 64; largest diameter 5-15 mm) were treated in vitro with piezoelectric shockwaves using the Edap LT-01 lithotripter with 2.5 Hz at either 100% or 54% power. The number of fragments larger than 2 mm was counted after every 30 seconds. The stones were defined as totally broken when all fragments were < 2 mm. Total fragmentation time was correlated with the energy level and the size of the stone. The number of large fragments did not correlate with the energy level but rather with the original size of the stone.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.