Abstract
Objective: To investigate the influence of the stimulus parameters on perimetry at various distances and draw conclusions for the clinical exploration of nonorganic visual loss. Methods: Visual field testing using Goldmann kinetic perimetry was performed on 15 healthy volunteers. The I/1e isopter at 33 cm was compared to the I/1e, II/1e and I/2e isopters at 66 cm. The 0/1e isopter at 33 cm was compared to the 0/1e, I/1e and 0/2e isopters at 66 cm. Results: Doubling the examination distance without adjusting the stimulus parameters resulted in significant perimetric visual field constriction. Doubling the stimulus diameter resulted in perimetric visual field expansion by a factor of 2.26 and 3.32 for I/1e and 0/1e, respectively. Increasing stimulus luminance by a factor of 3.17 caused expansion by a factor of 2.15 and 2.32 for I/1e and 0/1e, respectively. Conclusions: To avoid falsely diagnosing visual field constriction, stimulus parameters need to be adjusted when visual field testing is performed at double distance. Increasing stimulus luminance was more appropriate than augmenting stimulus size.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.