Abstract
EN ISO 11137 established regulations for setting or substantiating the dose for achieving the desired sterility assurance level.The validation studies can be designed in particular for different types of products. Each product needs distinct protocols for bioburden determination and sterility testing.The Microbiological Laboratory from Irradiation Processing Center (IRASM) deals with different types of products, mainly for the VDmax25 method. When it comes to microbiological evaluation the most challenging was cotton gauze. A special situation for establishing the sterilization validation method appears in cases of cotton packed in large quantities. The VDmax25 method cannot be applied for items with average bioburden more than 1000CFU/pack, irrespective of the weight of the package. This is a method limitation and implies increased costs for the manufacturer when choosing other methods.For microbiological tests, culture condition should be selected in both cases of the bioburden and sterility testing. Details about choosing criteria are given.