Abstract

For surgical and medical tools to prevent the spreading of infectious microorganisms to patients, disinfection and sterilization must be achieved by the use of disinfectants and sterilization procedures. Health care regulations must determine whether cleaning, disinfection, or sterilization is essential based largely on the item’s intended use because sterilizing all patient care equipment is unnecessary. In this work, successful sterilization of vials and rubbers used for medical purposes has been implemented. It can be claimed that the sequence of moist and dry heat is more practical for sterilizing primary packaging materials used in diagnostic or treatment facilities, especially for borosilicate vials and chlorobutyl, bromobutyl rubbers. It has been shown that the suggested sterilization method provides sterile and endotoxin-free vials and rubbers that can be used for medical purposes. Endotoxins and sterility tests have been implemented for microbiological analysis of borosilicate vials and chlorobutyl, and bromobutyl rubbers. Endotoxin levels in all vials and rubbers tested were within acceptable limits, and no microbial growth was observed in test specimens based on sterility tests. The suggested sterilization method was effective in producing sterile and endotoxin-free vials and rubbers suitable for medical use.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call