Abstract
To compare the accuracy of stereotactic localization using magnetic resonance imaging (Siemens 1.5-T Magnetom; Siemens, Erlangen, Germany) with two-dimensional and three-dimensional data acquisition techniques. A phantom study was performed in which the coordinates of an array of rods were determined from images in both two-dimensional and three-dimensional studies and compared with measured values in a series of transverse, coronal, and sagittal images. The results demonstrated a distinct advantage in using three-dimensional acquisition; an error greater than 2 mm was identified in only 0.8% of the imaged volume, compared with 12% of the imaged volume in the two-dimensional study. The results indicated that more accurate stereotactic localization is achieved with a three-dimensional acquisition.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.