Abstract

A phantom study has been carried out to assess the accuracy of stereotactic localisation, using magnetic resonance imaging. The stereotactic coordinates of an array of Perspex rods within the phantom were determined and compared with measured values, in a series of transverse, coronal and sagittal images. In the transverse plane, the maximum errors experienced were X = 2.3 mm and Y = 10.7 mm. If the third fiducial plate, at the front of the frame, were not used in the scaling of the images, there was considerable improvement in the Y direction (maximum error Y = 2.1 mm). However, some deterioration in the accuracy in the X direction resulted, particularly at the extremes of Z (maximum error X = 3.5 mm). In the coronal plane, the maximum errors were X = 1.8 mm and Z = 8.0 mm. With the third plate off, the errors decreased to X = 1.9 mm and Z = 3.3 mm. In the sagittal plane, the maximum errors recorded were Y = 1.1 mm and Z = 7.5 mm. It is not possible to calibrate in this plane without the third plate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.