Abstract

BackgroundModular total hip prostheses with ceramic heads are well established in orthopedic surgery and widely used. With the variety of different manufacturers and available designs, components are at risk for mismatch. Several case studies show the potentially devastating effects of mismatch. MethodsThe aim of this study was to investigate the outcome of one arbitrary component mismatch with commercially available components that appear to provide a stable fixation during assembly. A biomechanical in-vitro analysis of fracture strength (n=5) was carried out in accordance with ISO 7206-10. “Type1” Bi-Metric®-stems were mismatched with “V40” Al2O3 ceramic heads. FindingsMean fracture strength was reduced to about 50% of the recommended FDA minimum by the mismatch (Mean 23.68kN, SD 2.35kN). A small contact area between head and stem taper was identified as a potential key parameter. InterpretationMixing and matching components can put a patient at greater risk for ceramic head fracture and must be avoided.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.